EasyManua.ls Logo

WYLER BlueSYSTEM BASIC - Page 19

WYLER BlueSYSTEM BASIC
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
page 19 of 84
cont.
SENSOR
(BlueMETER
BASIC only)
- Possible error messages in the status line of the display:
"no sensor": no sensors found yet. NO sensors are connected via cables and no
instruments with wireless data transmission are registered.
In case the sensors should have been connected through cables, a REFRESH must
be performed after these have been connected. Press the key <ON/MODE>
repeatedly until [REFRRESH] appears on the display. Confirm with <ENTER>.
Now the measuring mode and the sensors can be selected. Confirm with <ENTER>
and select the required configuration using the <ZERO/SELECT ±> key.
measuring instrument port "A"
measuring instrument port "B"
measuring instrument port "A - measuring instrument port "B"
measuring instruments port "A" and "B" simultaneously
During the selection the following picture will be shown (example)
Confirm the configuration selected with <ENTER>. The address of the sensor "A"
can now be selected with the keys <ZERO/SELECT
±>. Confirm the selection with
<ENTER>.
Select the address of the first sensor (measuring instrument) using the keys
<ZERO/SELECT
±> and confirm with <ENTER>.
The second sensor (measuring instrument) can be selected with the same
procedure
When setting the measuring system in operation next time a selection as described
above is not necessary again. The configuration last used will automatically be
started.

Related product manuals