ABS.ZERO
(BlueMETER
BASIC only)
Setting of absolute ZERO with a reversal measurement.
Using the reversal measurement is a simple way to determine the exact ZERO OFFSET of the
instrument as well as the exact inclination of the surface where the instrument is placed.
Slide the instrument onto a flat, horizontally levelled surface (e.g. engineer's surface
plate).
The position of the instrument is to be marked on the surface.
Use the key <ON/MODE> to get to the menu point [ABS.ZERO] and confirm with
<ENTER>
On the BlueMETER BASIC the following display is seen:
Use the zapper or confirm with <ENTER>. On the display the progress of the
collected values are visible in graphic form.
When the curve is getting a straight line confirm the first value with the zapper or with
<ENTER>
After the first reading the following display is seen:
This display means the first reading was successful. The instrument must be turned
180 degrees without lifting and slid exactly onto the previously marked position
Use the zapper or confirm with <ENTER>. On the display the progress of the
collected values are again visible in graphic form.