EasyManua.ls Logo

WYLER BlueSYSTEM BASIC - Page 58

WYLER BlueSYSTEM BASIC
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
page 58 of 84
G SPECIAL FUNCTIONS
G1 R
ESET TO FACTORY PRE-SETTINGS
You can reset all adjustments and settings to the factory pre-settings. For this action press simultaneously the keys
<ENTER> and <ON/MODE> until an arrow to the right hand side appears in the display.
The following values will be set and the following actions performed:
BlueMETER:
• Filter Type 3
• Unit mm/m
• Display mode Absolute
• Relative Base in millimetres, value 1000
• Relative Base in Inch, value 10
• All members of the wireless data transmission group are deleted
• All members of the list of instruments are deleted
BlueLEVEL
• Filter Type 3
• Unit set to
o 1μ instrument: mm/m, 3 decimals
o 5μ instrument: mm/m, 3 decimals, rounded to 5µm/m
o 10μ instrument: mm/m, 2 decimals
• Display mode Absolute
• Relative Base in millimetre, value 1000
• Relative Base in Inch, value 10
• All members of the wireless data transmission group are deleted
• Relative Zero is set to 0
• Absolute Zero is set to 0
G2 FIRMWARE VERSION
With a special key operation you can read the version number of the firmware installed. After turning off the
instrument hold the <ON/MODE> key down for another 10 seconds.
The display shows in large figures the date of issue and at the bottom the version number of the firmware.