G SPECIAL FUNCTIONS
G1 R
ESET TO FACTORY PRE-SETTINGS
You can reset all adjustments and settings to the factory pre-settings. For this action press simultaneously the keys
<ENTER> and <ON/MODE> until an arrow to the right hand side appears in the display.
The following values will be set and the following actions performed:
BlueMETER:
Filter Type 3
Unit mm/m
Display mode Absolute
Relative Base in millimetres, value 1000
Relative Base in Inch, value 10
All members of the wireless data transmission group are deleted
All members of the list of instruments are deleted
BlueLEVEL
Filter Type 3
Unit set to
o 1μ instrument: mm/m, 3 decimals
o 5μ instrument: mm/m, 3 decimals, rounded to 5µm/m
o 10μ instrument: mm/m, 2 decimals
Display mode Absolute
Relative Base in millimetre, value 1000
Relative Base in Inch, value 10
All members of the wireless data transmission group are deleted
Relative Zero is set to 0
Absolute Zero is set to 0
G2 FIRMWARE VERSION
With a special key operation you can read the version number of the firmware installed. After turning off the
instrument hold the <ON/MODE> key down for another 10 seconds.
The display shows in large figures the date of issue and at the bottom the version number of the firmware.