H TECHNICAL DATA BLUESYSTEM BASIC
H1 T
ECHNICAL DATA OF THE RADIO MODULES BLUESYSTEM BASIC (OPTIONAL)
SENDER / RECEIVER
SENDER / EMPFÄNGER
BlueSYSTEM with
Bluetooth™ wireless technology
Frequency ISM-Band / 2,4000 - 2,4835 GHz
Modulation
FHSS
(Frequency Hopping Spread Spectrum)
Used Net-structure Point to point / Point to multi-point
RF Output power Max. +17 dBm / Class 1
Sensitive level Receiver -80 dBm
Batteries BlueLEVEL / BlueTC
Batteries BlueMETER
2 x 1.5V, Size “C” Alkaline
3 x 1.5V, Size “C” Alkaline
H2 T
ECHNICAL DATA BLUELEVEL BASIC
Sensitivity / Empfindlichkeit
1 µm/m
0.2 Arcsec
5 µm/m
1 Arcsec
10 µm/m
2 Arcsec
Display range external display /
Anzeigebereich externe Anzeige
± 10 mm/m ± 50 mm/m ± 100 mm/m
Limits of error / Fehlergrenze
<
0.5 Full-scale
(DIN 2276)
max. 1% of measured value /
max. 1% des aktuellen Messwertes
Limits of error / Fehlergrenze
>
0.5<Full-scale
(DIN 2276)
max. 1% of (2 x measured value - 0.5 x Full-scale) /
max. 1% von (2 x aktueller Messwert - 0.5 x Messbereichsendwert)
Temperature error / Temperaturkoeffizient /°C max. 0.1% Full-scale / max. 0.1% Messbereichsendwert
Display available / Anzeige verfügbar within 3 seconds / innerhalb von 3 Sekunden
Digital output / Digitalausgang
RS232 / RS422 / RS485, asynchron, 7 DataBits, 2 StopBits,
no parity, 9600 bps
External power supply
Externe Stromversorgung
BlueMETER BASIC:
+ 5V DC, max. 450 mW or/oder 8…28 V DC
Operating temperature range / Betriebstemperatur
Storage temperature range / Lagertemperatur
0 ... +40°C
-20 ... +70°C
Net weight without measuring base,
including batteries and handle
Netto-Gewicht ohne Messbasis,
inklusive Batterien und Griff
BlueLEVEL BASIC: 1200g
BlueMETER BASIC: 775g