Introducon6
Guidanceandmanufacturersdeclaraon–
electromagnecemissions
The ForeSite SS system is intended for use in the electromagnec
environments specied below. ForeSite SS system users should ensure
that it is used in such an environment.
Emissions Test Compliance Electromagnec
Environment Guidance
RF emissions CISPR 11 Group 1 The ForeSite SS system
uses RF energy only for
its internal funcon.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference
in the nearby electronic
equipment.
RF emissions CISPR 11 Class A The ForeSite SS system
is suitable for use in all
establishments other
than domesc and those
directly connected to the
public lowvoltage power
supply network that
supplies buildings used
for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuaons/
icker emissions
IEC 61000-3-3
Complies
Recommendedseparaondistancesbetween
portableandmobileRFcommunicaonsequipmentandtheForeSiteSS
The ForeSite SS is intended for use in an electromagnec environment in which
radiated RF disturbances are controlled. The customer or the user of the ForeSite SS
can help prevent electromagnec interference by maintaining a minimum distance
between portable and mobile RF communicaons equipment (transmiers) and the
ForeSite SS as recommended below, according to the maximum output power of the
communicaons equipment.
Rated maximum
output power
of transmier
W
Separaon distance according to frequency of transmier
M
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.24
0.1 0.37 0.37 0.74
1 1.17 1.17 2.34
10 3.69 3.69 7.38