ForeSite SS Hardware32
Cleaning
Cleaning may be performed if a part of the system becomes accidentally
dirty or smudged. It is strongly recommended that only health care
professionals aempt to perform any type of cleaning on any part of the
system, as follows:
Tablet Terminal
The Display Terminal may be wiped gently with a computer cleaning
cloth or clean towel to remove smudges from the touch screen.
Do not use liquid or aerosol cleaners, abrasive cleaners, waxes or
solvents as these may cause permanent damage to the screen.
NEVER immerse in water or other uids.
Power Cords/Connectors
No electrical device should come in contact with water or other
uids. This applies to the connector ends of USB cords and any other
electrical connector. A clean, dry cloth may be used to clean these
porons of the cords only, but ONLY while they are unplugged.
The non-connector porons of power cords and connectors may
be cleaned with a spray disinfectant. The following choices are
recommended:
ED Everyday Disinfectant by Wood Wyant, 4 ml/litre, 10 min.
contact
MetriGuard® by Metrex, 3 min. contact
Aer applying a light spray to non-electrical porons of power
cords, wait the recommended contact me for the parcular spray,
and wipe clean with a lightly damp, clean cloth. Allow to air dry
completely before aempng to plug in or use the cables again.
Sensor
If the sensor has one of the following
markings, it is cleanable as per
the Cleanable Sensor Instrucons
secon of this User Guide.
CLEANABLE
CLEAN ONLY AS
RECOMMENDED
IN USER GUIDE