ForeSite SS Hardware
33
For an UNSOILED Sensor:
→ NOTE: the sensor should never be immersed in water
6
H2O
4
1:10
1:10
3
cl
1 : 10
H2O
21
5
H
2
O
For sensors WITHOUT this, protecve sleeves are provided by XSENSOR
to shield against soiling. The sensor should not be exposed to moisture
and should never be immersed in water. In the case that something is
accidentally spilled on the sensor:
• Wipe carefully with a damp cloth and isopropyl alcohol if absolutely
necessary
• NEVER use a steam cleaner on the sensor
• Aer cleaning the sensor, allow to air dry COMPLETELY before
returning to use.
Cauon: Under no circumstances should a phenolic-based cleaning
soluon be used.
CleanableSensorInstrucons
1. Lay the unsoiled sensor on at surface
2. Prepare a 1:10 cleaning soluon (1
bleach : 10 water).
3. Dip a so cloth into 1:10 cleaning
soluon.
4. Clean sensor with so cloth moistened
with cleaning soluon.
5. Wet clean so cloth with tap water.
6. Clean residual soluon o with wet
cloth.