Introducon
9
Theequipmentisintendedforuseintheelectromagnecenvironment
speciedbelow.ThecustomerortheuseroftheEquipmentshould
assurethatitisusedinsuchanenvironment.
Immunity test IEC 60601 test
level
Compliance level Electromagnec
environment –
guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be
wood, concrete
or ceramic
le. If oors
are covered
with synthec
material, the
relave humidity
should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines ±1
kV for input/
output
lines
±2 kV for power
supply lines ±1
kV for input/
output lines
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Surge IEC 61000-
4-5
±1 kV line(s) to
line(s) ±2 kV
line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power
quality should be
that of a typical
commercial
or hospital
environment.