pg | 40
Sonde
Expansion
(Handheld)
Power
RS-232
SDI-12
+ 9-16V DC
- Ground
1 Amp Fuse
Voltage
esondeisdesignedtorunwith12-voltbatteries,with
supplyvoltagesbetween9and16VDC.
ese systems
should be directly powered by a sealed battery or installed
as part of a remote solar system.
PowertheSOAthrougha
1-ampslow-blowfuseforprotection.
Removethefuseuntil
allwiringiscompleted.

When used with a solar system always ensure use of
12-volt solar regulator. Never connect the sonde directly to
solar panel; voltages above 16.5 volts will cause the sonde to
shut down and excessive voltage will permanently damage
the sonde and is not covered under warranty.