EasyManua.ls Logo

Xylem EXO Series - Calibrate Menu

Xylem EXO Series
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pg | 54
KOR Software
Calibrate Menu
is menu is the main interface for calibrating EXO sensors. Calibration and verication settings will vary
by sensor. A device must be connected to access the Calibrate menu.
For detailed calibration procedures for individual sensors, reference Section 5 Calibration.
Calibration menu overview
In the Calibrate menu, the list of installed sensors is shown
on the le side of the window. If your sonde is congured
with a depth sensor, Port D depth also appears in the list.
Sensor calibration menu
Select a sensor from the list to bring up a sensor-specic
menu of parameters. For example, selecting ODO (optical
dissolved oxygen) from the list, brings up a menu for
the enabled parameters ODO % sat and ODO mg/L. (To
change the enabled parameters, go to the Options | Units
menu.)
Click on a parameter to bring up the sensor-specic
calibration menu. e manufacturer recommended default
calibration parameters are in bold.
A typical calibration window shows 1, 2, or 3 calibration
points, depending on the sensor. If the sensor supports
only a single calibration point, then the other calibration
points are not active (grayed out).
Temperature: is eld displays current temperature, if a
conductivity/temperature sensor is installed. If no sensor is
installed, user can input data into this eld.
Barometer: Displays for DO calibrations only.
Standard Value: User-input eld for the calibration
setpoint based on the value of the standard being used.
4.3

Related product manuals