IM 01E20F02-01E
5-8
5. EXPLANATION OF BASIC ITEMS
5.6 Integral LCD Indicator
Employing 32*132 full dot matrix backlit LCD,
various display can be obtained.
5.6.1 Flow Data Display
By the transducer block parameters setting in
DISPLAY_SELECT1, 2, 3 as described in 5.3, up to
three lines display can be made among the following
data.
-Flow Rate(%)
-Flow Rate
-Flow Rate(Bar)
-Integrator1 Out
-Integrator2 Out
-Arithmetic Out
-PD Tag
-Adhesion Check
-Communication
The number of the data can be configured by setting
DISPLAY_SELECT1, 2, 3 as follows;
1 line display 2 line display 3 line display
Display Select1 - - -
Display Select2Off Other than "Off" Other than "Off"
Display Select3 - Off Other than "Off"
T0507.eps
The data titles are displayed together with the flow
data and units.
F R
I T 1
I T 2
T A G
A D H
C O M
: Flow Rate, Flow Rate(%)
: Integrator1 Out
: Integrator2 Out
: Arithmetic Out
: PD Tag
: Adhesion Check
: Communication
A R
F0504.eps
Flow Rate (%)
1
0
0
.
0
%
FR
F0505.eps
Decimal point is always to the first place.
Flow Rate
Flow rate is displayed together with the units set in
XD_SCALE, the maximum number of figures is six.
-
1
3
.
4
5
6
k
g
a
l
/
i
n
m
FR
F0506.eps
In the case of L_TYPE is "Direct " decimal point
location of the data becomes as follows.
XD_SCALE.Decimal_Point
XD_SCALE.Decimal_Point
Desimal
Desimal
Point
Point
Location
Location
Example
Example
00123
11123.4
22
3
123.45
3
123.456
Other Auto
Refer to below
T0508.eps
When XD_SCALE.Units_Index is set as Other decimal
point location of the data becomes as follows.
XD_SCALE.EU_0,EU_100
XD_SCALE.EU_0,EU_100
set
set
value
value
Desimal Point
Desimal Point
Location
Location
Example
Example
XD_SCALE value -10000
0-33333
-10000 < XD_SCALE value -1000
1 -4444.1
-1000 < XD_SCALE value -100 2 -555.12
-100 < XD_SCALE value -10 3 -66.123
-10 < XD_SCALE value < 10 3 -7.123
0.123
10 XD_SCALE value < 100 3 66.123
100 XD_SCALE value < 1000 2 555.12
1000 XD_SCALE value < 10000 1 4444.1
10000 XD_SCALE value 0 33333
T0509.eps
*XD_SCALEvalue : Max (|XD_SCALE.EU0|, |XD_SCALE.EU100|)
When the flow data exceeds the maximum number of
figures six, then "84:Disp Over Wng" message is
displayed.
9 9 9
.
9 9
l / h
9
2 3 4 5 6 M l1 IT1
F R
pD8 4 : i s n gr WeO v
F0507.eps

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXF Series and is the answer not in the manual?

YOKOGAWA ADMAG AXF Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXF Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals