7-1
7. DEVICE INFORMATION
7. DEVICE INFORMATION
7.1 DEVICE STATUS
Device status for the AXF are indicated by using parameter DEVICE_STATUS_1 to DEVICE_STATUS_7 (index
1045 to 1052) in Resource Block.
Table 7.1 Contents of DEVICE_STATUS_1 (Index 1045)
T0701.EPS
0 0x00000001 Link Obj.16/32 not open
1 0x00000002 Link Obj.15/31 not open
2 0x00000004 Link Obj.14/30 not open
3 0x00000008 Link Obj.13/29 not open
4 0x00000010 Link Obj.12/28 not open
5 0x00000020 Link Obj.11/27 not open
6 0x00000040 Link Obj.10/26 not open
7 0x00000080 Link Obj.9/25 not open
8 0x00000100 Link Obj.8/24/40 not open
9 0x00000200 Link Obj.7/23/39 not open
10 0x00000400 Link Obj.6/22/38 not open
11 0x00000800 Link Obj.5/21/37 not open
12 0x00001000 Link Obj.4/20/36 not open
13 0x00002000 Link Obj.3/19/35 not open
14 0x00004000 Link Obj.2/18/34 not open
15 0x00008000 Link Obj.1/17/33 not open
16 0x00010000
17 0x00020000
18 0x00040000
19 0x00080000 101:Comm EEPROM Fault Communication EEPROM(FB) failure
20 0x00100000
21 0x00200000
22 0x00400000 41:RS O/S Mode RS. MODE_BLK.Target is O/S mode
23 0x00800000 100:Sim. Jmpr On Sim.enable Jmpr On
24 0x01000000 21:DL Incomplete Download incomplete
25 0x02000000 22:Download Fail Download fail
26 0x04000000
27 0x08000000
28 0x10000000
29 0x20000000
30 0x40000000
31 0x80000000
bit Hex Indicator description
Table 7.2 Contents of DEVICE_STATUS_2 (Index 1046)
T0702.EPS
0 0x00000001 10:uP Fault Microprocessor (CPU) failure
1 0x00000002 11:EEPROM Fault EEPROM failure
2 0x00000004 12:A/D (H) Fault A/D converter failure
3 0x00000008 13:A/D (L) Fault A/D converter failure
4 0x00000010 14:A/D (Z) Fault A/D converter failure
5 0x00000020 15:Coil Open Flowtube coil is open-circuit
6 0x00000040 16:EEPROM Dflt EEPROM default values
7 0x00000080
8 0x00000100
9 0x00000200
10 0x00000400
11 0x00000800
12 0x00001000
13 0x00002000
14 0x00004000
15 0x00008000
16 0x00010000 100:Comm uP Fault Communication uP failure
17 0x00020000 101:Comm EEPROM Fault Communication EEPROM failure
18 0x00040000 102:IT1 Not Saved IT1 save error
19 0x00080000 103:IT2 Not Saved IT2 save error
20 0x00100000 104:Comm Error1 AXF internal communication error
21 0x00200000 105:Comm Error2 AXF internal communication error
22 0x00400000 106:DL Incomplete Download is not completed
23 0x00800000 107:Download Fail Download failure
24 0x01000000 108:Not Ready Function block not scheduled
25 0x02000000
26 0x04000000
27 0x08000000
28 0x10000000
29 0x20000000
30 0x40000000
31 0x80000000
bit Hex Indicator description

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXF Series and is the answer not in the manual?

YOKOGAWA ADMAG AXF Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXF Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals