6-6
IM AQ6370C-17EN
Standard Event Status Register
Read
This register can be read by the common *ESR? query.
Its contents will be cleared when read. The read action complies with the IEEE 488.2
standard.
Write
Contents of the register can be cleared. The register can be cleared but not written to.
Clear
This register will be cleared under any of the following conditions.
Common *CLS command
Common *ESR? query
The clear action complies with the IEEE 488.2 standard.
Standard Event Status Enable Register
Read
This register can be read by the common *ESE? query.
The read action complies with the IEEE 488.2 standard.
Write
This register can be written by the common *ESE command.
The write action complies with the IEEE 488.2 standard.
Clear
This register will be cleared under any of the following conditions.
Data “0” is set using the common *ESE command.
Power ON
The register cannot be cleared in the following cases.
Receipt of the *RST command
Receipt of the *CLS command
Device clear (DCL, SDC)
The clear action complies with the IEEE 488.2 standard.
6.3 Standard Event Status Registers

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ6370C and is the answer not in the manual?

YOKOGAWA AQ6370C Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ6370C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals