<7. Errors and Countermeasures (Display unit)>
67
IM 01E24A01-01EN
Warning
The device works normally and measurement is also normal but warning occurs.
NE107
Status
Error Message
Error Description Countermeasure
Display
S 080:AO 1 saturate Saturation of Analog output 1 was detected. Check process value and current output 1 parameter setting.
S 081:AO 2 saturate Saturation of Analog output 2 was detected. Check process value and current output 2 parameter setting.
S 082:Pls 1 saturate Saturation of Pulse output 1 was detected. Check process value and pulse output 1 parameter setting.
S 083:Pls 2 saturate Saturation of Pulse output 2 was detected. Check process value and pulse output 2 parameter setting.
S 084:AI saturate Saturation of Analog input was detected. Check current input value and current input parameter setting.
C 085:Cable miscon Misconnection of cable was detected. Check the signal cable and excitation cable connection.
C 086:Coil insulation Insulation deterioration of coil was detected. Contact Yokogawa service center.
M 087:Adhesion lv 3
The resistance value of electrode exceeded Level 3.
(Adhesion detection of insulation to electrode)
Recommend cleaning electrode.
N 088:LC warn Decrease of conductivity was detected. Check uid conductivity.
M 089:Insu detect Insulation deterioration of electrode was detected. Contact Yokogawa service center.
C 092:AZ warn
Result of Autozero adjustment |Mag Flow Zero| exceeded
10 cm/s.
Check uid is stopped when executing auto-zero adjustment.
C 093:Verif warn Interruption of verication function was detected. Execute Verication again.
C 094:Fact noise warn Fluctuation of ow became larger. Check the uid.
C 095:Simulate active
Test mode was executed for any of Flow velocity,
Volumetric ow rate, Mass ow rate, Calorie, Current
output, Pulse output, Status input, Status output.
Release simulation or test mode.
S 096:AO 1 x It was detected that xied value is set to Current output 1. Check current output 1 is test mode or not.
S 097:AO 2 x It was detected that xied value is set to Current output 2. Check current output 2 is test mode or not.
S 098:Pls 1 x It was detected that xied value is set to Pulse output 1. Check pulse output 1 is test mode or not.
S 099:Pls 2 x It was detected that xied value is set to Pulse output 2. Check pulse output 2 is test mode or not.
S 100:AI x It was detected that xied value is set to Current input. Check current input is test mode or not.
C 101:Param restore run Restore function of parameter is running.
N 102:Disp over Number of digits available for display exceeded the limit. Check Display format parameter.
N 103:SD size warn Free space of microSD card decreased to less than 10%. microSD card may run out of memory space.
M 104:Bkup incmplt Parameter backup failed. Retry parameter backup.
S 105:SD mismatch Mismatch of microSD card was detected. Change microSD card.
M 106:SD removal ERR Removal of microSD card failed. Remove microSD in appropriate rocedure.
N 131:Trans mismatch Mismatch of sensor and transmitter was detected. Contact Yokogawa service center.
Information
The device works normally and measurement is also normal. Just reference information.
NE107
Status
Error Message
Error Description Countermeasure
Display
N 120:Watchdog Error of Watchdog timer was detected. Contact Yokogawa service center.
N 121:Power off Power-off was detected.
N 122:Inst power FAIL Instantaneous power failure was detected.
N 123:Param bkup run Parameter backup is running.
N 124:Data log run Data log is running.
N 130:DevID not enter No data entry of Device ID was detected. Contact Yokogawa service center.
N 132:Modbus cfgupdate
The Modbus conguration has been changed, but the
device is not operating according to its conguration.
Restart Modbus communication.
N 133:G/A mismatch
As a result of self-test, inconsistency of internal circuit was
detected.
Contact Yokogawa service center.
N 246:Simulate SW OFF Writing to parameters is locked.
Change the RB Write Lock (RB. WRITE_LOCK) to Not Locked or
turn off the hardware write lock switch, depending on RB Feature
Selection (RB.FEATURE_SEL).
N 248:Simulate SW ON Software or hardware simulation switch is ON.
Delete the value of MTB Sim Enable Message (MTB.SIM_ENABLE_
MSG) or turn off the hardware simulation switch.
Errors and Countermeasures (Display unit)
7

Other manuals for YOKOGAWA AXW4A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AXW4A and is the answer not in the manual?

YOKOGAWA AXW4A Specifications

General IconGeneral
BrandYOKOGAWA
ModelAXW4A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals