1-8
IM 701610-61E
Explanation
I
2
C Trigger Type
You can select from the following three types.
START (Start Trigger)
When a Start condition is detected, a trigger is activated on the first falling edge of the
SCL signal.
...
...
SDA
SCL
Start condition
1
2
3
4
5
6
7
8
9
Acknowledge bit
Data byte
Trigger activated here
1
2
3
4
5
6
7
8
9
Stop condition
Address + R/W bit
Non-ACK (Non-Ack Trigger)
When the Acknowledge bit is not present (when the SDA signal is set to “H”), a trigger
is activated on the 9
th
falling edge of the SCL (clock) signal.
Address + R/W bit
...
...
SDA
SCL
Start condition
1
2
3
4
5
6
7
8
9
Acknowledge bit
Data byte
Trigger activated here
1
2
3
4
5
6
7
8
9
Acknowledge present (L)
Acknowledge not present (H)
Stop condition
The Acknowledge bit of the start byte and Hs mode master code is not applicable.
ADRS (Address Trigger)
When the specified address and data pattern match, a trigger is activated on the 9
th
falling edge of the SCL (clock) signal.
Address + R/W bit
...
...
SDA
SCL
Start condition
1
2
3
4
5
6
7
8
9
Acknowledge bit
Data byte
Trigger activated here
1
2
3
4
5
6
7
8
9
Stop condition
1.3 Setting the Trigger Conditions