EasyManua.ls Logo

YOKOGAWA DL1640

YOKOGAWA DL1640
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-11
IM 701610-61E
1
I
2
C-Bus Signal Analysis Function
I2C on Pattern
This setting is used when activating the trigger based on the SCL/SDA signal and
CH3/CH4 pattern. The trigger is activated only when the I
2
C bus trigger condition is
met while the CH3/CH4 trigger condition is met.
CH3=L, CH4=H
891 189
I
2
C bus condition met
Condition not met
Trigger does not activate
SDA
SCL
CH3
CH4
I
2
C bus condition met
..........
.....
.....
Condition met
Trigger activated here
.....
Setting the CH3 and CH4 Pattern
H: The trigger source level is above the preset trigger level.
L: The trigger source level is below the preset trigger level.
X: Not used as the trigger source.
Setting the Trigger Condition
True: Activates a trigger when the specified CH3/CH4 pattern is met.
False: Activates a trigger when the specified CH3/CH4 pattern is no longer met.
I2C Pattern
This setting is used when activating the trigger based on the SCL/SDA signal and a
preset CH3/CH4 pattern. The SCL/SDA trigger met condition is held until the CH3/
CH4 trigger condition is met.
CH3=L, CH4=H, Enter
891 189
SDA
SCL
CH3
CH4
I
2
C bus condition met
.....
Condition met
Trigger activated here
Condition not met
Trigger does not activate
Holds the I
2
C bus
trigger met condition
.......... .....
.....
1.3 Setting the Trigger Conditions