Check the Physical Appearance
Do not use the instrument if there is a problem with its
physical appearance.
During Insulation Resistance Measurement
A high voltage is present at the probes during insulation
resistance measurement. Do not touch the DUT or the
earth or line terminal.
Immediately After Insulation Resistance Measurement
The probes or the DUT may remain highly charged.
Do not touch them immediately after the completion of
measurement.
Measure Properly
• Donotapplyavoltagethatexceeds600V(voltageto
ground) to the instrument.
• Donotapplyavoltageexceedingtheupperlimit.
Measurement Category
ThisinstrumentisratedtoCATIII600V.Donotusethe
instrument in locations that exceeds this category.
Probes
• UsethededicatedprobessuppliedbyYokogawafor
this instrument.
• Donotuseprobesthathavedeterioratedorare
defective.
• RemovetheprobesfromtheDUTbeforeattaching/
detaching the probes to/from the instrument.
Case Insulation
Donotusetheinstrumentifithasbeensubjectto
strong shock as a result of dropping it or hitting it against
anotherobject.Theshockcandestroytheprotective
insulation. Contact your nearest YOKOGAWA dealer for
repairs.