24
IM MY600-01EN
During Insulation Resistance Measurement
Rated voltage 50 V 100 V 125 V 250 V 500 V 1000 V
Range (auto range) 4/40/
100 MΩ
4/40/
200 MΩ
4/40/
250 MΩ
4/40/
400/500
4/40/400/
2000 MΩ
4/40/400/
4000 MΩ/
Display range
4 MΩ: 0.000 to 4.199 MΩ
40 MΩ: 3.60 to 41.99 MΩ
100 MΩ:
36.0 to
104.9
200 MΩ:
36.0 to
209.9
250 MΩ:
36.0 to
262.4
400 MΩ: 36.0 to 419.9 MΩ
500 MΩ:
360 to
524 MΩ
2000 MΩ:
360 to
2099 MΩ
4000 MΩ:
360 to
4199 MΩ
Center value 2 MΩ 5 MΩ 5 MΩ 10 MΩ 100 MΩ 200 MΩ
Over-range display >104.9
>209.9
>262.4
>524
>
2099
>
4199
Open-circuit voltage
100 to 110 % of rated measurement voltage
Short-circuit current
Within 1.5 mA
Rated measurement
current
1.0 to 1.1 mA
0.05 MΩ 0.1 MΩ 0.125
0.25 MΩ 0.5 MΩ 1 MΩ
First effective
measurement
range and accuracy
(tolerance)
0.100 to
10.00
0.100 to
20.00
0.100 to
25.00
0.100 to
50.0 MΩ
0.100 to
500 MΩ
0.100
to
1
000 MΩ
±2%rdg±2dgt
Second effective
measurement
range and accuracy
(tolerance)
10.01
to
100.0
20.01
to
200.0
25.01
to
250.0
50.1
to
500 MΩ
501
to
2000 MΩ
1001
to
4000 MΩ
±5%rdg
0.050
to
0.099 MΩ: ±2% ±4dgt
Other measuring
ranges and accuracy
(tolerance)
0.000
to
0.049 MΩ: ±2%rdg ±6dgt
Maximum capacitive load: 1 μF
The maximum capacitive load that can be discharged within
a specified period (10 s) after measurement (IEC61010-2-
034).
Measurement capacitive load: 2 μF
Capacitive load that is kept within ±10% fluctuation in the
JISC1302 output voltage test.