26
IM MY600-01EN
Item Specifications
Auto power-off The instrument turns off automatically, after
a beep sound, if there is no user interaction
for about 10 minutes (except when a
measurement is in progress).
LCD backlight, LED
light
Turns off automatically if there is no user
interaction for about 2 minutes (except when a
measurement is in progress).
Dimensions 97 (L) x 156 (W) x 46 (H) mm
Weight Approx. 490 g (including batteries)
Power source Four AA batteries (AA alkaline batteries
recommended)
Operating Environment and Conditions
This instrument complies with the EMC standard under
specific operating environment and operating conditions.
If the installation, wiring, and so on are not appropriate,
the compliance conditions of the EMC standard may not
be met. In such cases, the user will be required to take
appropriate measures.
Operating Uncertainty
Operating uncertainty (B) is an error obtained under the
nominal operating conditions and calculated with the
intrinsic error (A), which is an error of the instrument used,
and the error (En) due to variations. According to the
IEC61557 and JIS C 1302, the maximum operating error
should be within ±30 %. Intrinsic error (A) is uncertainty
of the performance characteristics under the reference
condition.