Model EXA PH202S-F & PH202S-P
No revision to drawing without prior
FM Approval
Title : FM Control Drawing PH202S-F & PH202S-P (Intrinsic safe Entity
concept)
Number : FF1-PH202S-00 Page : 8 of 10
Revision : 5.4
YOKOGAWA EUROPE B.V.
Date : 01/07/2004
〈 Sensor(s) are of a passive type to be regarded as ’simple apparatus’, devices which
neither store nor generate voltages over 1. 5 V, currents over 0.1 A, power over 25 mW or
energy over 20 µJ, or are FM Approvals entity approved and meet connection
requirements.
〈 Electrical data of the EXA PH202S-F & PH202S-P:
- Supply circuit:
Maximum input voltage Vmax=24 V
Maximum input current Imax=250 mA
Maximum input power Pi=1.2 W
Effective internal capacitance Ci=737 pF; Effective internal inductance Li=2.6
µH.
- Sensor input circuit:
Maximum output voltage Vt=14.4 V; Maximum output current It= 32.3 mA
Maximum allowed external capacitance Ca=600 nF
Maximum allowed external inductance La= 36 mH
〈 Any FM Approved barrier may be used that meets the following requirements:
Voc or Vt
≤ 24 V
Ioc or It ≤ 250 mA
Poc or Pt
≤ 1.2 W
Ca
? 737 pF + Ccable; La ? 2.6 µH + Lcable
When installing this equipment, follow the manufacturer s installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 Installation of Intrinsically
Safe Systems for Hazardous (Classified) Locations and the National Electrical Code
(ANSI/NFPA 70).
Associated apparatus connected to the barrier must not use or generate more than
250 Vrms
or Vdc.
〈 Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0
Ohm.
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power
before servicing or read, understand and adhere to the manufacturer s live
maintenance procedures.
Unclassified Location
Classified Location
Division 1
FM Approved
barrier
Voc (Vt) ¡Ü24 V
Ioc (It)
¡Ü 250 mA
Poc (Pt)
¡Ü 1,2 W
Ca
¡Y 737pF+ Ccable
La
¡Y 2,6 H + Lcable
I.S.
certified
Terminator
EXA
PH202S-F
& PH202S-P
Sensor
Connections
Max. cablelength: 60 mtr.
Cable dia. : 3 1 2 mm.
FM Class I, DIV. 1, Group ABCD
T3B for ambient temp. ≤ 55 ϒC
T4 for ambient tem