6-10
IM WT5000-01EN
To prevent noise-related errors, the trigger feature has a hysteresis of approximately 1% when the crest
factor is set to CF3. For example, when the trigger slope is set to
, a trigger will occur if the input signal
level falls approximately 1% below the trigger level and then passes through the trigger level on a rising
edge. The trigger feature has a hysteresis of approximately 2% when the crest factor is set to CF6 or
CF6A.
The trigger feature for displaying waveforms does not work in the following circumstances.
When the update mode is set to constant interval
During integration
Therefore, the waveform display start point (the signal level at the left edge of the screen) may not be
stable. Also, the numeric data measurement period may not be synchronized with the waveform data
measurement period.