152 Operation
G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0
Printed on 09. 10. 2006
Checklist
Warning!
If a function fails, you must not use this instrument for safety reasons. Cor-
rect the fault (see the "Troubleshooting table“) or contact our service dept.
Always check the following points before surgery (without patient!):
Check that the correct rated voltage has been set.
Check that all cables have been connected.
Check that the light guide has been connected.
Turn on the system at the power switch of the suspension system.
VISU 160 surgical microscope
Check the zoom function
by pressing the appropriate button on the foot control panel.
Check the focusing function
by pressing the appropriate button on the foot control panel.
Friction adjustment of the surgical microscope
Check that the friction of the surgical microscope's rotation has been
adjusted as required using the friction adjustment knob on the suspen-
sion system.
Speeds of the surgical microscope's functions
Check that the speeds of the microscope's functions have been ad-
justed as required on the suspension system.
Eyepieces / binocular tube
Check that the surgical microscope and the tube are in a position con-
venient for you.
Check that the correct interpupillary distance has been set.
Check that the eyecups have been adjusted in such a way that you
can see the full field of view.
Check that the correct prescription has been set on the diopter scale.
Check that image quality is the same throughout the entire magnifica-
tion range.
Checking the accessories
Using the manuals provided, check that the other equipment (illumination
system, video system, etc.) is functioning properly.