180 Operation
G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0
Printed on 09. 10. 2006
Failure of lamp control
• Press the Manual key (1) if brightness control is no longer possible.
Note:
When the manual function has been activated, all electrical control sys-
tems are inoperative. The lamp brightness is automatically adjusted to a
fixed setting.
Failure of focusing system
• Press Manual key (1) if, for example, the focusing system always
moves into its upper or lower end position.
• Focus by moving the suspension arm of the ceiling mount (floor stand)
up or down.