180 Operation
G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0
Printed on 09. 10. 2006
Failure of lamp control
Press the Manual key (1) if brightness control is no longer possible.
Note:
When the manual function has been activated, all electrical control sys-
tems are inoperative. The lamp brightness is automatically adjusted to a
fixed setting.
Failure of focusing system
Press Manual key (1) if, for example, the focusing system always
moves into its upper or lower end position.
Focus by moving the suspension arm of the ceiling mount (floor stand)
up or down.