154 Operation
G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0
Printed on 09. 10. 2006
If operated for 8 hours/day on 5 days/week, the lamp needs to be ex-
changed after 12 weeks maximum.
Balance setting
Check that the suspension arm has been properly balanced. When
the release button on the surgical microscope is pressed, the effort re-
quired to move the arm up or down must be the same.
Limitation of downward travel
The minimum working distance (height) from the surgical field has
been set using the adjustment screw for limiting downward travel.
S88 floor stand only: Stand base
Check that the brake tab has been pressed and that the stand is se-
curely locked in position.
Check the accessories
Using the manuals provided, check that the other equipment (surgical
microscope, coobservation tube, video system, etc.) is functioning
properly.
Foot control panel
Check that the plug of the foot control panel has been connected.
Check that the power switch of suspension system has been switched
on.
Check that all functions assigned to the respective keys on the foot
control panel are working properly.