G-30-1529-en OPMIĀ® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0
Printed on 09. 10. 2006
Maintenance / Further information 187
Suspension systems
Surgical field illumination
too bright.
Brightness level set too
high.
Adjust brightness using the
control on the suspension
system or the foot control
panel.
--
Switch off illumination on
suspension system.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
--
Focusing system inopera-
tive.
Focusing system does not
work, or moves to upper or
lower end position.
Use suspension arm to set
correct focal plane.
Contact service dept.
--
Zoom system inoperative. Zoom system does not
work, or moves to upper or
lower end position.
Manually adjust magnifica-
tion using zoom knob.
Contact service dept.
--
Zoom and focusing systems
inoperative.
Functions not correctly set
on suspension system.
Contact service dept. --
Microscope motion too stiff. Friction adjusting knob on
suspension system tight-
ened too firmly.
Slightly loosen friction ad-
justing knob.
page 124
DeepView inoperative DeepView is on mechanical
end stop.
Press the DeepView button
to deactivate the function.
--
Problem Possible cause Remedy See
Problem Possible cause Remedy See
No function at all. Line power failure. Contact in-house electri-
cian.
-
Power switch of suspension
system not switched on.
Press power switch. -
Automatic circuit breaker in
power switch of suspension
system has been activated.
Press power switch again. -