PROCEDURES
7.4
TEST
PROCEDURE
WARNING
e
Limit
internal
discharges
to
100
Joules
or
less.
Do
not
discharge
the
PD™
1400-series
device
internally
more
than
3
times
in
one
minute.
Multiple
internal
discharges
at
more
than
100
Joules
maydamage
the
unit
.
«
All
persons
near
the
equipment
must
be
warned
to
“STAND
CLEAR?”
prior
to
defibrillator
discharge.
e
Press
both
DISCHARGE
buttons.
The
unit
should
discharge
and
the
"1007
RDY"
message
should
be
removed
from
the
display.
"TEST
OK"
should
appear
on
CRT
and
stripchart
recording.
Mark
the
Checkout
Procedure
Data
Sheet.
11-19