R SERIES — NON-INVASIVE BLOOD PRESSURE (NIBP) OPTION INSERT
24 www.zoll.com 9650-0914-01 Rev. C
1. Select the proper size cuff, as described in “Selecting the Proper Size Cuff” on page 8.
2. Connect the hose to the R Series unit and to the cuff, as described in “Connecting the Hose”
on page 9.
3. Apply the cuff to a human test subject, as described in “Applying the Cuff to the Patient” on
page 10.
4. Set the cuff inflation pressure (if current setting is not appropriate), as described in “Setting
Cuff Inflation Pressure” on page 14.
5. Take a blood pressure measurement.
During cuff inflation, inspect cuff, hoses, and connectors for air leakage. If a leak is present,
correct, replace, or service item.
6.
Verify that the unit displays no error messages.
7. Verify that the displayed blood pressure values are correct.
8. With alarms enabled, verify that the patient alarms are functional by adjusting the high and
low limits and repeating the above blood pressure measurement. Check that the unit:
• emits a continuous audio tone, and
• highlights the alarming parameter’s value and flashes the alarm symbol on the display.
9. Visually inspect hose, hose connector, and cuffs for signs of damage. If damaged, replace the
damaged item.
Cleaning the Hose and Reusable Cuffs
You can clean the hose and reusable cuffs by wiping the surface with a damp cloth and
disinfecting with a mild disinfectant solution. Blow dry air through the hose before use;
likewise, ensure the reusable cuff is dry before use.
NIBP Accessories
The following accessories can be used with the R Series NIBP option:
NIBP CUFFS
Cuffs
Limb
Circumference REF
Thigh
(Reusable)
38 to 50 cm
(14.96 to 19.69 in.)
8000-1654
Large Adult
(Reusable)
31 to 40 cm
(12.20 to 15.75 in.)
8000-1653
Adult
(Reusable)
23 to 33 cm
(9.06 to 12.99 in.)
8000-1651
Small Adult
(Reusable)
17 to 25 cm
(6.69 to 9.84 in.)
8000-1650
Child
(Reusable)
12 to 19 cm
(4.72 to 7.48 in.)
8000-1655
Neonate #5
(Disposable)
8.0 cm to 15.0 cm
(3.1 to 5.9 in.)
8000-0644