R SERIES — NON-INVASIVE BLOOD PRESSURE (NIBP) OPTION INSERT
18 www.zoll.com 9650-0914-01 Rev. C
WARNING! If the R Series unit takes a measurement but detects the presence of artifact in the
signal (denoted by “*” in the NIBP display area), the measurement may not be
accurate. Under such circumstances, perform additional blood pressure measurements.
If you repeatedly obtain artifact, use alternate techniques to determine blood pressure
prior to taking clinical action.
When reading the blood pressure values on the display, keep in mind that the following
conditions can influence NIBP measurements:
Position of the patient
Position of the cuff relative to the patient’s heart
Physical condition of the patient
Patient limb movements
Convulsions or tremors
Very low pulse volumes
Premature ventricular beats
Vibrations in the cuff or hose caused by moving vehicles
Improper cuff size or application
WARNING! Do not begin NIBP measurements unless the patient type setting is appropriate for the
patient. Taking NIBP measurements on a pediatric or neonatal patient while the unit is
in adult mode can result in inaccurate measurements and injury to the patient. Taking
NIBP measurements on an adult patient while in pediatric or neonatal mode can result
in inaccurate measurements.
Taking a Single Measurement
You can take a single measurement by pressing the NIBP button ( ) unless the R Series
defibrillator is charged or charging.
Caution Make sure that no measurement is being taken when you start a new measurement. If you press
the NIBP button while the R Series unit is taking a measurement, the measurement aborts
immediately and the cuff deflates.
Caution Before taking the first measurement, a period of at least 5 minutes should elapse to give time for
blood pressure stabilization.

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB