R SERIES — NON-INVASIVE BLOOD PRESSURE (NIBP) OPTION INSERT
28 www.zoll.com 9650-0914-01 Rev. C
Specifications
Principle of Operation Oscillometric
Warm-up Time Operational in less than 10 seconds
Regulatory Standards ANSI/AAMI SP10-2002;
EN 1060-1: 1996 Specification for Non-invasive sphygmomanometers;
EN 1060-3: 1997;
EN 60601-2-30: 2000
NIBP Measurement Range
Systolic Adult: 40 – 260 mmHg; 5.3 – 34.6 kPa
Pediatric: 40 – 160 mmHg; 5.3 – 21.3 kPa
Neonatal: 40 – 130 mmHg; 5.3 – 17.3 kPa
Diastolic Adult: 20 – 200 mmHg; 2.7 – 26.7 kPa
Pediatric: 20 – 120 mmHg; 2.7 – 16.0 kPa
Neonatal: 20 – 100 mmHg; 2.7 – 13.3 kPa
Mean Adult: 30 – 220 mmHg; 4.0 – 29.3 kPa
Pediatric: 30– 133 mmHg; 4.0 – 17.7 kPa
Neonatal: 30 – 110 mmHg; 4.0 – 14.6 kPa
NIBP Measurement Accuracy Per AAMI SP10 +/- 5 mmHg (0.7 kPa) mean difference;
8 mmHg (1.1 kPa) standard deviation
a
NIBP Measurement Resolution 1 mmHg; 0.13 kPa
NIBP Measurement Cycle Time
Typical: 30 seconds
Worst Case: 180 seconds
Pulse Rate Range
30 – 220 pulses per minute
Pulse Rate Accuracy
±2% or ±3 bpm, whichever is greater
Pulse Rate Resolution
1 bpm (beats per minute)
Alarm Limits Systolic Adult High 80 – 260 mmHg; High 10.7 – 34.7 kPa
Low 40 – 140 mmHg; Low 5.3 – 18.7 kPa
Pediatric High 80 – 160 mmHg; High 10.7 – 21.3 kPa
Low 35 – 140 mmHg; Low 4.7 – 18.7 kPa
Neonatal High 60 – 130 mmHg; High 8.0 – 17.3 kPa
Low 25 – 120 mmHg; Low 3.3 – 16.0 kPa
Diastolic Adult High 50 – 200 mmHg; High 6.7 – 26.7 kPa
Low 25 – 100 mmHg; Low 3.3 – 13.3 kPa
Pediatric High 50 – 130 mmHg; High 6.7 – 17.3 kPa
Low 20 – 100 mmHg; Low 2.7 – 13.3 kPa
Neonatal High 30 – 105 mmHg; High 4.0 – 14.0 kPa
Low 20 – 100 mmHg; Low 2.7 – 13.3 kPa
Mean Adult High 60 – 220 mmHg; High 8.0 – 29.3 kPa
Low 30 – 120 mmHg; Low 4.0 – 16.0 kPa
Pediatric High 60 – 140 mmHg; High 8.0 – 18.7 kPa
Low 30 – 120 mmHg; Low 4.0 – 16.0 kPa
Neonatal High 35 –110 mmHg; High 4.7 – 14.7 kPa
Low 30 – 105 mmHg; Low 4.0 – 14.0 kPa