Cuff and Hose Safety
9650-0914-01 Rev. C R Series Non-Invasive Blood Pressure (NIBP) 3
Do not use damaged cuffs, hoses, or connectors.
Ensure proper cuff selection and placement to avoid inaccurate measurements or patient
injury.
Position the cuff so it is level with the heart during measurement.
Do not attach the cuff to a limb being used for IV infusion, SpO
2
monitoring, or other
monitoring equipment. Cuff inflation might block the infusion, causing harm to the patient
or inaccurate SpO
2
measurements.
Do not repeat NIBP measurements (particularly STAT measurements) at intervals less than
3-5 minutes over an extended period of time. Rapidly repeating measurement can impair
circulation in the monitored limb.
Do not sterilize or immerse the cuffs or hoses.

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB