Basic Operating Methods
65
2) Click the button in the upper right of the home page, option boxes
[Shutdown], [Log out], [Restart] will pop up;
3) Click [Log out] and [OK];
4) Back to software log in surface, re-log in with old account and password or a new
one.
3.17.Restart
1) Confirm that the system is in “Standby” state;
2) Click the button in the upper right of the software interface, option boxes
[Shutdown], [Log out], [Restart] will pop up;
3) Click [Restart] and [OK];
4) Back to software log in surface, re-log in with old account and password or a new
one.
3.18.Operation after shutdown
1) Confirm that the system is in standby state;
2) Click the button in the upper right of the software interface, option boxes
[Shutdown], [Log out], [Restart] will pop up;
3) Click [Shutdown] and [OK];
4) The software will be closed;
5) When the software is closed, you need to wait another 20 seconds after the screen
is off, so that the system is really shutdown;
6) Turn off the power switch to power off the analyzer.
Attention
When power off the analyzer, the reagent refrigeration system will
stop working, reagent kits should be taken out to store at 2~8 .
3.19.Operation after shutdown and power off
1) When shutdown the software and the power is still on;
2) Open the sample tray and reagent tray, take out the samples, calibrators,
controls and reagent kits;