Basic Operating Methods
41
3.4.Analyzer start up
3.4.1.Turn on your analyzer
Before inserting the power plug, check whether the main power switch of your
analyzer is in the state. If not, it is necessary to switch to the state first, and then
safely insert the power plug.
After properly inserting the power plug, turn the main power switch to make it in the “
state, the indicator lights up, the system starts up, and then enters the login interface.
3.4.2.Login
1) Enter the user name and password in the [Login] dialog box and click [OK].
Caution
1) The username of the system administrator is “Admin” by default,
and the password is “123456” by default. It is recommended that
the password should be changed during the first use to prevent
others from arbitrarily exploiting the administrator’s rights;
2) If the operator password is forgotten, contact the technical
service engineers of Zybio or its local agents.
2) Log in correctly and the software home page is displayed to complete the startup
process of your analyzer.
Caution
To ensure accurate test results, start testing operations at least 30
minutes after startup to ensure a stable light source and temperature.
3.5.Confirm your analyzer status
After your analyzer is started, confirm the various states of your analyzer if necessary.
For example: Reagent status, maintenance alarm status. When your analyzer status is
abnormal, refer to “Maintenance and Service” and “Alarm and Troubleshooting” for system
maintenance and troubleshooting.
3.5.1.Confirm reagent status
1) Click the [Reagent] button on the home page to check whether the residual