2) The pop-up box will push the minimum free sample position and barcode input
box by default;
3) Scan the barcode of the sample with an external barcode scanner;
4) After the barcode is successfully displayed, click [Acquire] to automatically obtain
the sample information and test items that need to be tested from the LIS;
5) A notification box of successful acquisition pops up and click [OK].
1) For sample barcodes, it is recommended to use Codabar, Code128,
EAN-13, Code 93, Code 39 (regular) barcode system, please
ensure that the keyboard switch to English lowercase input method
when inputting;
2) To print meaningful characters, it is recommended to use numbers
(0-9) and capital letters (A-Z);
3) Notice: If the English letter size switch of the barcode appears,
please bring up the keyboard to switch the upper and lower case
and scan the barcode of the sample tube again.
◼ Delete
1) Select [Request] - [Sample List];
2) Select a piece of sample information to be deleted in [Sample List];
3) Click [Delete], the delete confirmation prompt box will pop up, [Delete Selected
Sample] is default.
⚫ [Delete Selected Sample]: Delete selected sample;
⚫ [Batch Deletion]: Enter the start and end positions, delete samples in batch.
4) Click [OK] to delete.
◼ Scan samples automatically
Click [Scan Samples], the analyzer can scan the sample tray automatically, sample
information will be showed in [Sample List], when connect to LIS.
◼ Recall
Select sample tests that has been submitted but has not been executed, click
[Recall] to recall the test.
◼ Rotor
Click [Rotor], the sample position selection box will pop up, select the position from
the drop-down window and click [OK], and wait for the tray to rotate.
Note:
1) The deletion operation only deletes the sample from the [Request List], not
delete the sample result; The information editing and result viewing of