39
AD-8127/AD-8121B format “info 1” General format “info 2”
~ : Space, ASCII 20h
CR
: Carriage return, ASCII 0Dh
LF
: Line feed, ASCII 04Ah
Outputting the “Title block” and “End block”
When weight values are recorded as the GLP report, the “Title block” and “End block”
are added at the beginning and at the end of a group of weight values.
“Title block”
1. Press and hold the PRINT key.
Release the PRINT key when
5tart is displayed. The
balance outputs the “Title block”.
2. Press the PRINT key or select
the auto-print mode to output the
weight values.
___________A_&_D
MODEL_____EJ-123
S/N____Q12345678
ID________ABCDEF
DATE__2013/11/14
TIME____02:53:21
CAL.TEST(EXT.)
ACTUAL
________0.000__g
_____+100.001__g
TARGET
_____+100.000__g
SIGNATURE
-_-_-_-_-_-_-_-_
~~~~~~~~~~~
A
~
&
~
D
<CRLF>
MODEL~~~~~EJ-123
<CRLF>
S/N~~~~Q12345678
<CRLF>
ID~~~~~~~~ABCDEF
<CRLF>
DATE
<CRLF>
<CRLF>
TIME
<CRLF>
<CRLF>
CAL.TEST(EXT.)
<CRLF>
ACTUAL
<CRLF>
~~~~~~~~0.000~~g
<CRLF>
~~~~~+100.001~~g
<CRLF>
TARGET
<CRLF>
~~~~~+100.000~~g
<CRLF>
SIGNATURE
<CRLF>
<CRLF>
<CRLF>
----------------
<CRLF>
<CRLF>
<CRLF>
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero value
Target weight value
Target weight
Signature
•
•
•
•
5tart
0.000
To output “Title block”
12.345
23.456
To output weight values
Press and hold
end