GB 27
English
Calibration report example for weighing sensor compliant with GLP, GMP and ISO
For the AD-8121 printer, use printer MODE 3.
11.3. Calibration of Drying Temperature for MX-50
The temperature calibrator (accessory AX-MX-43)
adjusts the drying temperature on the pan.
Put the sensor on the pan and input measurement
data at 100°C and 160°C.
Each adjustment needs fifteen minutes.
The buzzer sounds at the end.
t-Up is displayed after no adjustment for five
minutes during the operation and calibration is
stopped. Press any key to return to weighing
mode.
Refer to the instruction manual of the certified
temperature calibrator (accessory AX-MX-43).
CALIBRATED WEIGHT
CAL.WEIGHT 20.000 g
SIGNATURE
- - - - - - - -
A&D
MODEL
S/N
ID
DATE
TIME
MX-50K
K1234567
LAB-123
2002/04/15
13:57:24
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration mass
Signature
Sensor pan
Thermometer