AND MX-50/MF-50 Instruction Manual
GB 6
The moisture analyser MX-50 and MF-50 are built using a super hybrid sensor (S.H.S.) originaly
designed for high resolution laboratory balances. Therefore, the result is more precise and reliable.
An analyser with S.H.S. needs a sample mass of only a few grams and this makes the measurement
time shorter.
A 400W halogen lamp is used as the heating source and the temperature on the sample pan may
reach 200°C in as quick as two minutes.
The Standard Mode (Refer to section 7.1) can easily provide the moisture content at high accuracy
automaticaly using pre-set conditions.
There are three User Modes that the user can quickly set their own measurement parameters.
Automatic mode When the Drying Rate (change of moisture content per unit of time %/s) becomes
less than the preset value, the measurement will be automatically completed.
Timer mode The sample is dried for a preset time period.
Manual mode This user can terminate the measurement by pressing the STOP key.
The analyser can store and recall individual measurement settings. Maximum numbers of setting
stored in the analyser differs by model.
Maximum number MX-50 20 sets
MF-50 10 sets
The Data Memory can store measured results and can transfer the data to PC or a printer at a later
time. Maximum numbers of data possible, stored in the analyser, differs by model.
Maximum number MX-50 results of 100
MF-50 results of 50
The software "WinCT-Moisture", an accessory of the MX-50, has 3 separate software programs
– RS Key can allow data from the moisture balance to be placed into any Windows based program.
– RS Com can allow the remote control of the analyser from a PC and will record data into it’s own
text sheet.
– RS Fig can graphically show the change of moisture content and other information to a connected
PC in realtime.
The software "WinCT", an accessory of the MF-50, has 2 software programs RS Key and RS Com as
described above.
The analyser is equipped with a serial interface as standard for connection to a printer or a PC.
The calibration of the weighing part of the analyser can be adjusted using calibrated weights.
The temperature calibration of the MX-50 can be adjusted using the dedicated temperature calibrator
(AX-MX-43).
The analyser can output GLP, GMP and ISO compliant data to a printer at the end of either calibration
process.
The analyser has a self test function that can detect functional errors.
The analyser displays the Drying Rate, [%/ min], during measurement. It can be used as a reference to
find suitable analysing modes.
Sample pans can be used repeatedly if traces of the previous sample are completely removed.
A calibration sample of Sodium tartrate dihydrate is supplied as a standard accessory to check the
function and accuracy of the moisture analyser.
A reference card can be found underneath the analyser for quick reference.