AND MX-50/MF-50 Instruction Manual
GB 36
Specifications
14
MX-50 MF-50
Measurement method 400 W Halogen lamp
Drying temperature range at sample pan
50 °C to 200 °C (1 °C increments)
Temperature calibration By Accessory AX-MX-43 ---
Sample weight range 0.1 g to 51 g
Accuracy: Repeatability of measurement, (Standard deviation)
Moisture over 5 g sample 0.02% 0.05%
content *
1
over 1 g sample 0.1% 0.2%
Weighing mode 0.001g 0.002g
Minimum reading
Moisture content 0.01%, 0.1% 0.05%, 0.1%, 1%
Weighing mode 0.001g 0.002g
Measurement programs
Analysing Standard mode Set accuracy from HI, MID or LO and select drying temperature.
mode Sample weight and termination value is set automatically.
When the drying rate reaches fixed termination value,
measurement is automatically completed. (*
2
)
Automatic When drying rate is less than preset termination value,
complete mode
measurement is automatically completed. (*
2
)
Timer mode After heating at the preset drying time, measurement is
automatically stopped. (1min. to 480 min.)
Manual mode When pressing the key at any time, measurement is stopped
and the result is decided.
Measurement unit Moisture content (Wet-base)
Moisture content (Dry-base, Atro)
Solid content
Ratio (%)
Number of memory 20 set 10 set
Data memory function Storage of 100 results Storage of 50 results
Communication function RS-232C serial interface
Application software for Windows (CD-ROM)
WinCT-Moisture Analysing utility
WinCT Communication software
Operation environment
5 °C to 40 °C (41°F to 104°F), 85%RH or less (no condensation)
Sample pan 85 mm diameter
Power source, Maximum current (rms), AC 200 V to 240 V, 1.5A
Maximum consumption 50Hz or 60Hz, Approximately 400W
External dimensions 215(W) x 320(D) x 173(H)mm,
Mass (Net weight) Approximately 6kg (without accessories)
Housing material Flame-retardant (UL94V0)
*
1
: After preheating the analyser, the data can be obtained with approximately 5 g calibration sample
(Sodium tartrate dihydrate) in standard mode at 160 °C
*
2
: When the drying rate reaches to the preset termination value, the measurement is completed.
User mode