EasyManua.ls Logo

A&D MX-50 - Maintenance; 13.1.1. Replacement of the Halogen Lamp

A&D MX-50
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AND MX-50/MF-50 Instruction Manual
GB 32
Turn off the power switch and remove power cord during maintenance.
Cool down all parts of the analyser before maintenance.
Remove pan support, sample pan and breeze break ring.
Clean the analyser with a lint free cloth that is moistened with warm water and a mild detergent.
Do not use organic solvents to clean the analyser.
Dry the parts and reassemble them. Refer to " Section "5.1. Installing the Instrument"
Use the original packing material and box for transportation should the product need to be sent for repair.
Replace the halogen lamp, when the drying time is excessive or the lamp is defective. Use the
halogen lamp accessory AX-MX-34-240V. The life time of the halogen lamp is approximately 5000 hours.
13.1.1. Replacement of the Halogen Lamp
Caution
Clean the surface of the halogen lamp. If there is a stain or fingerprint, it may reduce the life
of the halogen lamp. Use gloves.
Fix the lamp wire to the hook so that the lamp wire does not touch the glass-housing and
heater cover.
1. Turn off the power switch and remove power cord.
2. Check that the lamp is cool.
4. Remove glass-housing with two screws.
5. Remove halogen lamp.
6. Install the new halogen lamp so that the light will shine
downwards.
7. Fix the lamp wire to the hook.
8. Fix the glass-housing with two screws.
Screw
Heater
Halogen lamp
Fix the wire
to the hook
Plug in connector
Light shines
downwards
Wire
Rated voltage label
Glass-housing
Maintenance
13

Other manuals for A&D MX-50

Related product manuals