ABB PGC5007 - Analyzer will Not Power up (X-Purge); Repeated Column Failures; Short Quartz Tube Life

ABB PGC5007
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PGC5007
PGC5007 Operating Instructions 4 Maintenance information
22 OI/PGC5007-EN, Rev C
Insufficient or
No Air
Power
Disconnected
Check circuit
breakers and local
disconnect switch
ANALYZER WILL
NOT POWER UP
(X PURGE)
Air Leak in Purged
Housing
X-Purge
Locked Out
Analyzer will not power up.
Leak check
with leak
detector
Cycle Power to reset
X-Purge and allow X-
Purge to sequence
SHORT QUARTZ
TUBE LIFE
Furnace Too Hot
Reduce furnace
operating temperature
Furnace is running too hot.
REPEATED COLUMN
FAILURES
Excessive Column Failures
Replace columns
Columns have broadened
out and lowered peaks.
Thermocouple not
In position