EasyManuals Logo

Accu-Chek Compact Plus Reference Manual

Accu-Chek Compact Plus
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #192 background imageLoading...
Page #192 background image
192
Symbols, error messages and troubleshooting
Error message Meaning
The test strip is stuck or the test strip
drum does not rotate.
Press the Test button to turn the meter off.
Open the drum compartment cover and (if
applicable) pull out the test strip. Press the
red drum release button and remove the
test strip drum. Insert the test strip drum
again and close the drum compartment
cover. The meter rotates the drum and
reads the bar code. When you next turn on
the meter to perform a test, the drum
symbol and flashing bottle symbol are dis-
played. Perform a performance check. You
may use all the test strips remaining in
the drum. The 90-day use-by period will
no longer be calculated for this drum. The
drum symbol will therefore be displayed
each time you perform a test until you
insert a new test strip drum.
AC_CP_LCM_Manual_EN-UK final.indd 192_CP_LCM_Manual_EN-UK final.indd 192 20.6.2007 11:04:4620.6.2007 11:04:46

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH