PSPIX² • User manual
32
reservation time for each work station is con gurable from the PSPIX acquisition module in the “Setup”
menu from the “Con guration” tab.
NOTE:
When the PSPIX system is reserved for a given workstation number, a noti cation will appear in the
acquisition module of the other users. It is impossible for any other workstation to reserve the PSPIX
system during this period.
2. Then follow steps 2 to 7 for the single user con guration described above for reading the IP.
3. To manually free up the PSPIX system click on the “Home”
icon in the bottom right-hand corner
of the PSPIX screen, OR click on the “locked” icon
on your acquisition module. You will return to the
PSPIX multi-user home page.
NOTE:
If an image cannot be transferred onto the computer due to a network error, computer failure or
software bug, the last image read is saved in the PSPIX system. A caution window appears on the Touch
screen of the PSPIX when no other IP can be inserted for reading.
You are offered three options:
- Attempt to re-establish the Ethernet connection with the computer to automatically transfer the image. To
do this ensure that the Ethernet cable is correctly connected OR ensure that the PSPIX acquisition module
is correctly open.
If the Ethernet connection is reconnected then a new window con rming the reconnection will be displayed
during the image transfer time.
- Delete the X-ray image from the memory by clicking on the “Trash” icon
.
Con rm the deletion by pressing the OK button.
-Collect the image from the PSPIX acquisition module of another computer connected to the PSPIX system (in
the case of a multi-user con guration). Take care to open the correct patient le in the imaging software,
then access the “Setup” menu, then the “Image” tab in the PSPIX acquisition module. Finally click on the
“Retrieve the Last Scanned Image” button.
CAUTION:
Risk of PSPIX system malfunction. Do not put an IP back in the insertion slot of the PSPIX system until the
previous IP has fallen into the IP receptacle.