DIP SWITCH ON OFF PARAMETER
1 INSTALLED NOT INSTALLED TUBEHEAD 1
2 dc ac TUBEHEAD TYPE 1
3 INSTALLED NOT INSTALLED TUBEHEAD 2
4 dc ac TUBEHEAD TYPE 2
5 INSTALLED NOT INSTALLED II° CONTROL BUTTON
6 LONG 12’’ = 30cm SHORT 8’’ = 20cm CONE
7 NOT AVAILABLE
8 NOT AVAILABLE
INSTRUCTION
To change the number of the tubehead installed, move the dip-switch n° 1 or the dip-switch n° 3:
DIP SWITCH 1 3
ON ON
OFF OFF
►if the tubehead is connected to the XRAY1 terminal board, put the dip-switch n° 1 in the ON position; otherwise
in OFF position;
►if the tubehead is connected to the XRAY2 terminal board, put the dip-switch n° 3 in the ON position; otherwise
in OFF position.
To change the type of tubehead installed, move the dip-switch n° 2 or the dip-switch n° 4:
DIP SWITCH 2 4
ON ON
OFF OFF
►if the tubehead connected to the XRAY1 is ac, put the dip-switch n° 2 in the OFF position; otherwise put it in
the ON position;
►if the tubehead connected to the XRAY2 is ac, put the dip-switch n° 4 in the OFF position; otherwise put it in
the ON position.
PLEASE NOTE
It is possibile to connect to the timer of type AC: n° 2 x-mind ac® tubeheads
To change the amount of control buttons installed, move the dip-switch n° 5:
DIP SWITCH 5
ON
OFF