INSTALLATION & MAINTENANCE MANUAL • X-MIND AC • V2 • (06) • 08/2016 • NXACEN020B
ENGLISH
8.3. ACCURACY OF THE EXPOSURE TIME
1. Put the X-ray detector at the cone outlet centred on the primary beam.
2. Turn on the timer.
3. Set the exposure time to 0.2s on the timer.
4. Carry out the exposure.
5. At the end of the exposure read on the display the actual exposure time.
6. Check the accuracy of the exposure time, use the chart below:
CONFORMITY CHECK CHART
read the time value on the display
Actual exposure time
read the measure on the instrument II°
Emission time
subtract I° from II° III°
divide III° by II° IV°
Absolute error
multiply IV° by 100
Percentage error (%)
the accurancy is checked when the % error is
± 5 % or ± 20msec
By adoptiong the most favorable condition
PLEASE NOTE
Otherwise, call the “Assistance Service”.