137
2. Start a new 2WIN (Refraction) measurement from KALEIDOS App. KALEIDOS App
imports the patient data from the CSV input file and displays the data. After the im-
port is completed, KALEIDOS App will delete the CSV input file from the Shared
Folder.
NOTE: KALEIDOS App is automatically checking the Shared Folder approximately
every 2 seconds for a new input file.
3. If the patient data displayed in KALEIDOS App are correct, start the measurement.
4. Save the exam. The CSV output file containing the exam results is automatically cre-
ated in the shared folder and it is retrieved by the EMR system. Then the EMR system
deletes the CSV output file.
5. Review the patient’s exam results on the EMR system.
A CSV input file triggers an output file. If your EMR system does not provide an input
file in the Shared Folder to begin with, your device will not write output files, and the
Shared Folder will remain empty.
Once a CSV input file is provided, the manual entry of patient data in the KALEIDOS
App user interface is disabled.
You can manage multiple 2WIN-S on the same EMR system. Different 2WIN-S
devices work on different Shared Folders, so you must create a dedicated Shared
Folder for each 2WIN-S.