Example configuration file
0
8
0
16
0
32
0
This file creates the following sequences of exam phases:
Phase1 = dark condition (no visible illumination), 2 seconds
Phase2 = 8% of the maximum visible light is turned on, 2 seconds
Phase3 = dark condition, 2 seconds
Phase4 = 16% of the maximum visible light is turned on, 2 seconds
Phase5 = dark condition, 2 seconds
Phase6 = 32% of the maximum visible light is turned on, 2 seconds
Phase7 = dark condition, 2 seconds
NOTE that any value greater than 100 value will be considered 100%, while 0 stands for
dark condition. The configuration file will not be considered if there will be a different number
of rows or different kind of characters, 2WIN will use the standard measurement mode in
this case.
Calibration mode
In this operating mode it is necessary to edit a file “pupilresponse.txt” and copy it in the SD
card of the 2WIN.
The operator must decide 2 parameters and then 4 intensities of light for the calibration of
the system in this specific mode.
The two parameters are: