8230 Optical Power Meter Operation Manual
A-3
A.2.1 Causes of the Generated Calibration Errors
Table A-3 Causes of the Generated Calibration Errors
No. Cause Command
Displayed
err code
Status Remarks
1 The IV offset calibration is not conver-
gent.
XIVC Err 10 ERR Bit7 ON
2 The difference between the VIG gener-
ation value, which is input when the
zero or full calibration is performed,
and the measurement value exceeds
50% of the full scale.
The measurement value exceeds the full
scale when calibrating.
XPC Err 10 ERR Bit7 ON
3 The full calibration is performed with-
out zero calibration.
XPC Err 15 ERR Bit13 ON Command execution error
4 The VIG generation value is not within
the input range.
XPC Err 15 ERR Bit13 ON Command execution error