MAINTENANCE MANUAL
IFR 6000
2-2-3
Page 22
Aug 1/06
STEP PROCEDURE
10. Set up the UUT for the first listed frequency below:
UUT Channel X ATC-1400 Frequency (MHz)
1 1025
40 1064
86 1110
126 1150
11. Press the RUN TEST Soft Key.
12. Verify the interrogation frequency displayed on the UUT equals Recorded
Frequency ± 20kHz.
13. Press the STOP TEST Soft Key.
14. Repeat the above 4 steps with each setting listed.