MAINTENANCE MANUAL
IFR 6000
2-2-3
Page 4
Aug 1/06
D. Verification
(1) Self Test
TEST EQUIPMENT: None
VERIFICATION FAILURE: If any step in this procedure fails or is out of tolerance,
this indicates a failure in the Test Set. Refer to
Troubleshooting for corrective action.
STEP PROCEDURE
1. Connect Test Set to appropriate AC power source with the External DC Power
Supply. Press Power Key to turn unit ON.
2. Press the SETUP Key repeatedly until the SETUP-GENERAL screen is
displayed.
3. Press the H/W TOOLS Soft Key to display the SETUP-HARDWARE TOOLS
screen.
4. Press the SELF TEST Soft Key to display the SETUP-SELF TEST Screen.
5. Press the RUN TEST Soft Key to initiate the Self Test.
6. Verify all tests pass.