Original Instructions Rev.1.0 User Manual Isa 11/ 42
Electrical & Environmental Requirements
The centrifuge requires must be 230V/50Hz or 115V/60Hz, single phase, (current rating varies per centrifuge model, refer
to § 1.4.1 Specifications) according to the manufacturer sheet on the back of the unit.
The centrifuge must be plugged into an outlet with protective earthing connection with the standard power cord.
The electrical outlet into which the centrifuge is connected should be readily accessible for maintenance purposes.
The power cord plug serves as the main disconnect device and should remain readily accessible. The electrical outlet should
be on its own branch circuit, protected with a circuit breaker at the distribution panel near the centrifuge.
The centrifuge is designed for use under the following ambient conditions (according to EN 61010-1)
• Indoor use
• Maximum altitude: 2000 m
• Ambient temperature range of 5 to 40 ° C
• Maximum relative humidity of 80% for temperatures up to 22 ° C,
• Supply voltage fluctuation <± 10% of rated voltage.
• Overvoltage on the Category II power supply network (IEC 60364-4-44 standard)
• Maximum degree of pollution of the room: 2.0
• Walls less than 1m away must be able to stop the centrifuge from moving in the event of a major accident.
Maximum performance is ensured in the following ambient temperature range: 15 ° C to 25 ° C.