Original Instructions Rev.1.0 User Manual – Isa 24/ 42
Use clear or distilled water to clean the rotor and buckets. A soft non-metallic brush or lint-free cloth can be used. Dry the
rotor and accessories ensuring the least accessible areas are wiped.
To ensure a proper swinging movement of the buckets,
clean all 8 trunions of the rotor.
Remove the leftover grease and dust.
Apply a thin layer of food quality grease (provided).
Also clean the buckets, more specifically the area in contact
with the trunions.
To prolong the life of accessories, you can then place a thin film of grease supplied with the centrifuge.
Rotors, buckets, inserts can be autoclaved at 122.5 ° C / 252 ° F / 215 kPa for 15 minutes
Note: When autoclaving, buckets, inserts, and rotors must be separated from each other.
Autoclaving causes accelerated aging of plastics, and may change the color of the plastic.
Perform regular cleaning to prevent any risk of corrosion in case of persistent impurity,
and to ensure the hygiene conditions of working equipment.
Bio hazardous, radioactive and/or toxic materials:
Before any intervention, to ensure the safety of the operator, please refer to the laboratory’s decontamination protocol.
The centrifuge and its accessories may be contaminated and require remediation in this case.